Description
Recommended for HPLC and GC use. Supplied with traceable calibration certificate and 2 Type K thermocouple probes. Features a built in printer and data-logging facility.
Specification:
2-channel digital thermometer with printing and logging facility, supplied with 2-type K thermocouples (suitable for low temperature use such measuring HPLC column oven temperature), and two rolls of chart paper. Specification - 0.1°C or °F resolution, Range -200 to +1370°C, accuracy 0.3% of reading, powered by 4 x 1.5V AA alkaline batteries. Traceable calibration of thermometer only at 5 setpoints: 0, 25, 100, 250 & 400°C with certificate.
Specification
Specification:
- Range and Accuracy (Type K ) -200 to 1370°C, ± (0.3% rdg +0.7°C), -328 to 2498°F, ± (0.3% rdg +1.4°F)
- Resolution: 0.1°C (in range of -200 to 650°C)
- Multi-point calibration (0, 25, 50, 100, 250 & 400°C.) with traceable calibration certificate
- Dual TC input (will only accept Type K)
- 2 x
VKIT-1516
Type K TC probes
- 4 x 1.5V AA batteries
- Supplied with sturdy carry case
Contents
Contents:
1 x VKIT-1531
2 x
VKIT-1516
type K thermocouples probes
1 x Calibration Certificate
2 x Thermal Paper Rolls
1 x Manual
4 x 1.5V AA batteries
1 x Carry case
Accessories
VKIT-1514
Place in the sample compartment tray or carousel to conveniently measure temperature.
VKIT-1517
Insert through the septum to quickly measure the temperature within the inlet.
VKIT-1519
Robust and easily fitted thermocouple probe, suitable for GC Operational Qualification. New and updated design.
VKIT-1518
Insert into the transfer line of Agilent 7697 and G1888 headspace samplers to easily measure temperature during OQ.
VKIT-1523
For accurate temperature measurement in FID detectors. Suitable for 6890, 7890 & 8890 GC.
VKIT-1532
10 x thermal chart rolls suitable for {product 11|link|code} printing thermometer.